SOP for HPLC Column Receipt, Checking and Regeneration

SOP for HPLC Column Receipt, Checking and Regeneration

Standard operating procedure to check the performance of the HPLC columns against the certificate received from vendor.

1.0 OBJECTIVE
To lay down the procedure for checking the performance of column with respect to the certificate received from the vendor and to carry out performance test as per in-house method. Also to describe the procedure for regeneration of column.

2.0 SCOPE
This procedure is applicable to all Reverse Phase (C8, C18, cyno, phenyl) column and Normal Phase (Si) Column regeneration.

3.0 RESPONSIBILITY
3.1 Doing: Technical Assistant, Executive
3.2 Checking: Executive/Manager

4.0 ACCOUNTABILITY
4.1 Head of the Department

5.0 PROCEDURE
5.1 Checking of new HPLC column
5.1.1 On receipt of a new column from the vendor, condition the column using methanol at             0.5 ml/min.
5.1.2 Check the column with respect to the certificate of analysis for back pressure,                       retention time, theoretical plates using the material provided with column and method           specified by the vendor. Record the results in Annexure-I
5.1.3 Carry out the column performance (Theoretical plates, Tailing, resolution and back                 pressure) using following chromatographic condition. Record the results in Annexure-I.
5.1.4 Performance of Reversed phase column.

Mobile phase: Methanol: water (70:30) v/v
Flow rate: 1.0ml/min.
Wavelength: 254nm
Inj.Volume: 20μl
Sensitivity: 0.005 AUFS

Test solution for C8 and C18: Mixture of Benzene and Toluene (0.1%) in the mobile phase.
Test solution for Cyano and Phenyl: Mixture of Benzene (0.1% ) and Naphthalene (4mg.)in 100ml mobile phase.

5.1.5 After checking, flush the column with methanol for 20 minutes at flow rate of                           1.0ml/min. in normal direction and store the column in methanol.
5.1.6 Performance of Normal Phase (Si) column

Mobile phase: IPA: Hexane (40:60) v/v
Flow rate: 1.0ml/min.
Wavelength: 254nm
Inj.Volume: 20μl
Sensitivity: 0.005 AUFS
Test solution: Mixture of Toluene (0.5 μl/ml) and Diethyl phthalate (2μl/ml) in mobile phase.

5.1.7 After checking, flush the column with Hexane for 20 min. at flow rate of 2.0ml/min and           store the column in Hexane.
5.2 Assign column number as follow.
      XXX YYY

Where XXX indicates the short form of the column, it may be 2 to 4 digit
e.g. ODS Column = ODS
BDS Column = BDS
Amine Column = NH
Silica Column = SI
Protein pack = PP
High mol. weight protein = HMWP
YYY indicates serial no. in series starting from 001.

5.2.1 Enter the details of the column in Annexure-I.
5.3 Column Regeneration for reverse phase column
5.3.1 Flush the column using warm water ( 55°C) at flow rate of 0.7 ml/min. for 10 minutes.           Maintain the column at (55°C) in column oven during flushing.
5.3.2 Inject 100 μl of DMSO four times at interval of 2 minutes.
5.3.4 Flush it with methanol at flow rate of 0.7 ml/min in reverse direction for 70 minutes.
5.3.5 After flushing with methanol, flush the column with chloroform for 70 minutes at flow             rate of 0.7 ml/min in reverse direction.
5.3.6 At the end, wash the column in methanol for 70 minutes at flow rate of 0.7 ml/min. in             reverse direction.
5.3.7 After regeneration of column, perform the column performance test as per Step 5.1.4             to 5.1.5
5.3.8 Store the column in methanol, after flushing with methanol for 20 minutes at flow rate           of 1.0ml/min. in normal direction.
5.3.9 Maintain the column History card on use of column during analysis as per Annexure-II.
5.4 Column Regeneration for Normal Phase Column
5.4.1 Flush the column with Hexane for 60 minutes at flow rate of 2.0 ml/min..
5.4.2 After flushing with Hexane, wash the column with Methylene chloride for 60 min at 
         flow rate of 2.0ml/min.
5.4.3 After flushing with methylene chloride, wash the column with Isopropyl alcohol for 60             minutes at flow rate of 2.0 ml/min.
5.4.4 At the end, flush the column with methanol for 60 minutes at flow rate of 2.0 ml/min.
5.4.5 After regeneration of column, perform the column performance test as per Step 5.1.6             to 5.1.7.
5.4.6 Store the column in Hexane, after flushing with Hexane for 20 minutes at flow rate of             2.0 ml/min in normal direction.
5.5 If the system suitability fails than only column regeneration required.

6.0 ABBREVIATIONS
6.1 ml = millilitre
6.2 min. = minute
6.3 °C = Degree centigrade
6.4 IPA = Isopropyl alcohol
6.5 μl = Microlitter
6.6 DMSO = Dimethylsulfoxide

ANNEXURE -I
HPLC COLUMN CHECK AND REGENERATION RECORD

Columns make :
Column receipt date :
Column Sr.no.(As per supplier) :
Column Id .No. :
Column inhouse no. :
Date of issue :
Column dimension :
Check of new column :

Sr. No.
Component
Dataasper COA
Observed Data


Plates
RT
Plates
RT
1





2





3





4





Back pressure bar







In-house check

No. of Theoretical Plates
Tailing
Resolution
Date of Performance check
Back pressure
Benzene Toluene
Toluene/Naphthalene Diethyl phthalate
Benzene Toluene
Toluene/Naphthalene Diethyl phthalate









                Performed By:                                                                                                                                                    Checked By:

Comments

Popular posts from this blog

Checklist for Audit in Sterile Area

Checklist for Audit in Sterile Area Checklist to verify the sterile manufacturing area before the audit. 1. Is there any SOP giving details for activity/ movement in the sterile area. 2. Verify the suitability of system for sanitization of hands. 3. Is there the SOP on entry-exit; gowning and de-gowning into the critical area available? 4. Verify the availability of following records:  Air pressure differential  Particulate monitoring  Temperature  Humidity  Microbiological monitoring by settle plate 5. Are these records available and maintained as per SOPs? 6. Check the above mentioned individual records. Are the entries within specifications mentioned in SOP/ MTPs. 7. Is the controlled copy of MTP pertaining to monitoring by settle plate available to plant person? 8. Are the location for settle plates identified and the chart/ layout available with plant? 9. Are the following areas monitored for parameters mentioned in step 4:  Air-lock to the critical ar...

Aseptic Filling Process (Media Fill) Validation Protocol in Sterile Pharmaceuticals

Aseptic Filling Process (Media Fill) Validation Protocol in Sterile Pharmaceuticals Validation of sterile manufacturing process by media fill validation test as per PICS guidelines for aseptic validation or aseptic process simulation. 1.0 OBJECTIVE 1.1 To define procedures for validating and maintaining the validation of all aseptic filling processes and qualification of the quality of the product by system/facility/equipment. 2.0 SCOPE 2.1 This procedure applies to all aseptically filled sterile products intended for human use. 3.0 RESPONSIBILITY 3.1 The Validation personnel coordinate the aseptic filling validation program and write the sterile media aseptic filling processes reports. 3.2 Manufacturing personnel the sterile media aseptic filling processes and performs the sterile media fills. 3.3 QA personnel perform the sampling and assist with the IPQA monitoring required for each sterile media fill. 3.4 QC personnel to perform the testing and assist with the monitoring required fo...

MSDS (Material Safety Data Sheet)

MSDS   What Is an MSDS? An MSDS (Material Safety Data Sheet) is a document that provides detailed information about a chemical substance, including its hazards, safe handling procedures, storage requirements, and emergency measures. Today, MSDS is more commonly called SDS (Safety Data Sheet) under the Globally Harmonized System (GHS). Why MSDS / SDS Is Important Protect workers and employees Ensure safe handling and storage of chemicals Prevent accidents and injuries Provide emergency response information Comply with workplace safety regulations.  MSDS (Material Safety Data Sheet)  International safety standards.  GHS – Globally Harmonized System Full form: Globally Harmonized System of Classification and Labelling of Chemicals Develop by United Nations (UN).  OSHA (USA) – Hazard Communication Standard (HCS) Factories Act & Chemical Rules (India) Factories Act, 1948 Manufacture, Storage and Import of Hazardous Chemical Rules, 1989 REACH (Europe) REACH Regula...

Steps for HPLC Method Development

Steps for HPLC Method Development Know about the different steps of the HPLC analytical method development in pharmaceutical analysis. Analytical method development is considered as a critical process in pharmaceuticals. Availability of the different types of columns, operating  parameters, mobile phase composition, diluent and pH values make it critical to develop an analytical method. A good analytical method should be  simple, used column, mobile phase and buffer should be common. It can be done easily step by step. Following are the common HPLC method development steps. 1. Selection of HPLC Analytical Method 2. Selection of Chromatographic Conditions 3. Parameter Optimization 1. Selection of HPLC Analytical Method: First of all consult the literature that is available on the product. It will help you to understand the nature of  the product that will help to select the different parameters. A. Sample Preparation: Select method to prepare the sample according to its ...

Blow Fill Seal (BFS) and Form Fill Seal (FFS) Technology in Sterile Manufacturing

Blow Fill Seal (BFS) and Form Fill Seal (FFS) Technology in Sterile Manufacturing Know about sterile pharmaceutical production using Blow Fill Seal (BFS) and Form Fill Seal (FFS) technology. Both of these techniques are used to manufacture  sterile pharmaceutical products   as parenteral (LVP & SVP), infusions, ophthalmic and inhalation products. These are automated techniques to prepare sterile products. The basic concept of the  FFS   and BFS is to reduce the  contamination   by forming the container, filling and sealing in a closed sterile chamber of the machine. There is no personnel intervention to reduce the chances of the contamination during the manufacturing of sterile products. It gives more production at very low operational cost with the high assurance of  sterility . Blow fill seal technology is widely used and accepted by the various pharmaceutical regulatory authorities as US-FDA and MHRA. The system is being used for over 30 years a...