Guideline for Preparation of Mobile Phase Required for HPLC
Learn guidelines for preparation and
validation of the mobile phase for HPLC analysis.
1. Use HPLC grade/ AR Grade / GR grade chemicals.
2. Weigh / Measure each component of the mobile phase
separately.
3. Degas each solvent and buffer by sonication and then
filter each component of the mobile phase separately through 0.45μ filter (Use
the filterpaper as per suitability with solvent).
4. Mix the individual solvent and buffer as per the
specified proportion.
5. Adjust the pH of the mobile phase wherever specified
(i.e. before or after mixing of individual components) between ± 0.05 unless
otherwise specified.
6. Sonicate the mobile phase for 10 minutes or more as required.
7. Ensure that mobile phase is free of particles. Do not use
the mobile phase if particles are observed.
8. Enter the details of mobile phase preparation in
laboratory note book.
Related: Theoretical Plates and their Determination in
HPLC Analysis.
Procedure for
mobile phase validation
1. Prepare the mobile phase as per the test procedure and
keep on bench top in well closed condition.
2. Evaluate system suitability parameter(s) as per the test
procedure at initial, after one day and after two days.
3. The mobile phase is considered stable if all the system
suitability parameter are within the specified limits.
Frequency: Mobile phase is to be prepared daily if mobile
phase validation is not done. In case if sequence is required to be run in
continuation for more than 24 hours & sufficient mobile phase is available
in the reservoir then sequence can be allowed to run by demonstrating system
suitability of analysis by injecting bracketing standard.
For stability validated mobile phase – As per frequency
indicated.