Skip to main content

Aseptic Filling Process (Media Fill) Validation Protocol in Sterile Pharmaceuticals

Aseptic Filling Process (Media Fill) Validation Protocol in Sterile Pharmaceuticals

Validation of sterile manufacturing process by media fill validation test as per PICS guidelines for aseptic validation or aseptic process simulation.

1.0 OBJECTIVE

1.1 To define procedures for validating and maintaining the validation of all aseptic filling processes and qualification of the quality of the product by system/facility/equipment.

2.0 SCOPE

2.1 This procedure applies to all aseptically filled sterile products intended for human use.

3.0 RESPONSIBILITY

3.1 The Validation personnel coordinate the aseptic filling validation program and write the sterile media aseptic filling processes reports.

3.2 Manufacturing personnel the sterile media aseptic filling processes and performs the sterile media fills.

3.3 QA personnel perform the sampling and assist with the IPQA monitoring required for each sterile media fill.

3.4 QC personnel to perform the testing and assist with the monitoring required for each sterile media fill.

4.0 ACCOUNTABILITY

4.1 Manager QA

5.0 REFERENCE (S)

5.1 In- house.

5.2 Pharmaceutical Inspection convention Guideline.(PIC)

6.0 PROCEDURE

6.1 EXIT AND ENTRY PROCEDURE

6.1.1 All Operators and Supervisors enter department as per Defined Gowning procedure.

6.2 AMPOULE FILLING MEDIA

6.2.1 The media fill run shall be performed for approximately the same duration as required for filling of a normal production batch (approx. 12 hours).

6.2.2 Batch (Media) manufacturing shall be done in the initial hours of the 1st shifts, followed by other processing steps normally followed.

6.2.3 Nitrogen flushing shall not to be done during media filling.

6.2.4 The filling of media shall be done in the following sequence:-

6.2.4.1 The media (Soybean Casien Digest Madium) batch shall be manufactured in the manufacturing area. The bulk is then aseptically filtered in to a sterile holding tank (under LAF) in the sterile area. The tank containing the filtered solution is then connected sterile silicon pipeline and filling done in ampule without employing the use of on-line cartridge filters/filter housing on the machine.

6.2.4.2 Fill volume of media as per the protocol. (The fill volume of media in units need not equal the fill volume of finished product units, the quantity filled shall be sufficient to wet all the inner surfaces of the unit & to enable an inspection to detect positives.)

6.2.4.3 Filling activity is performed in the filling room (especially the filling zone) under LAF.

6.2.4.4 Consecutive first three runs will be validated with sterile media solution.

6.2.4.5 Simulate the adverse condition testing of the system (details of adverse conditions given separately in step 6.4) after filling approximately 10,000 units.

6.2.4.6 Filling to be planned in a manner such that the operator covers the activities during the shift change over.

6.3 VIAL FILLING MEDIA:

6.3.1 The media fill run shall be performed for approximately the same duration as required for filling of a normal production batch (approx. 12hours).

6.3.2 Batch (media) manufacturing shall be done in the initial hours of the 1st shifts, followed by other processing steps normally followed.

6.3.3 Nitrogen flushing shall not be done during media filling.

6.3.4 The filling of the media to be done in the following sequence-

6.3.4.1 The media (Soybean Casien Digest Madium) batch is manufactured in the manufacturing area. The bulk is then aseptically filtered into asterile holding tank (under LAF) in the sterile area. The tank containing the filtered solution is then connected with is sterile silicon pipeline and filling done in vials without employing the use of on-line cartridge filters/filter housing on the machine and then dosing of sterile lactose in vials.

6.3.4.2 Transfer the sterile lactose through material airlock into the sterile area and transfer the sterile lactose aseptically into the previously sterilized hopper of the vial filling machine. Three runs will be validated with 250 mg sterile lactose and 5 ml media solution, 500 mg sterile lactose with 5 ml media solution and 1000 mg sterile lactose with 10 ml media solution to validate the weight range from 250 mg to 1000 mg.

6.3.4.3 Fill volume of media or weight of Powder as per the protocol. (The fill volume of media filled units need not equal the fill volume of finished product units, the quantity filled shall be sufficient to wet all the inner surfaces of the unit & to enable an inspection to detect positives.)

6.3.5 Filling activity is performed in the filling room (especially the filling zone) under LAF.

6.3.6 Simulate the adverse condition testing of the system (details of adverse conditions given separately in step 6.4) after filling approximately 10,000 units.

6.3.7 Filling to be planned in a manner such that the operator covers the activities during the shift change over.

6.4 ADVERSE CONDITIONS: (AMPULE/ VIAL)

6.4.1 Filling machine OFF: For 60 minutes (The filtered air supply in the filling zone not to be put off. Imitate maintenance activities that are likely to take place during routine filling process, after restarting approx. 1,000 units to be filled)

6.4.2 LAF Filtration OFF: For 5 minutes (filling to be done when LAF is OFF)

6.4.3 Air Handling Units power OFF: For 5 minutes (ALL the AHU’s of the sterile and the adjoining areas to be switched OFF, filling to be done when AHU’s are OFF).

6.4.4 Total Power OFF for 2 minutes: (ALL the AHU’s of the sterile and the adjoining areas, filling m/c, LAF’s etc to be switched OFF, after restarting approx. 1,000 units to be filled)

6.4.5 Highest (250 ampule /vial min.) speed of machine: approx. 1,000 units to be filled

6.4.6 Change in tubing’sApprox. 1,000 units to be filled

6.4.7 Ampule/Vial Sterilizing tunnel OFF: For 5 minutes (total power supply of the tunnel is to be switched OFF, after restarting tunnel all the units from the cooling zone to be filled).

6.5 STORAGE OF FILLING UNITS (AMPULE/VIAL)

6.5.1 Collect all units in a tray or suitable container, Store in previously maintained temperature defined area or incubator.

6.5.2 First Store at temperature 20 -25 ° C for seven days.

6.5.3 After that Store at temperature 30 -35 ° C for seven days.

6.6 RESULT

1. After the incubation period of the media-filled containers, they are visually examined for microbial growth. Contaminated containers should be examined for evidence of container/closure damage which might compromise the integrity of the packaging system. Damaged containers should not be included as failures (positives) when evaluating results.

2. The number of containers used for media fills should be sufficient to enable a valid evaluation. For small batches, the number of containers for media fills should at least equal the size of the product batch. The target should be zero growth and the following should apply:

• When filling fewer than 5000 units, no contaminated units should be detected.

• When filling 5,000 to 10,000 units,

a) One (1) contaminated unit should result in an investigation, including consideration of a repeat media fill;

b) Two (2) contaminated units are considered cause for re-validation, following an investigation.

• When filling more than 10,000 units:

a) One (1) contaminated unit should result in an investigation;

b) Two (2) contaminated units are considered cause for re-validation, following an investigation.

Abstracted from: PIC 007-5 dated 01 July 2009

Recommendation on the Validation of aseptic processes.

6.7 DESTRUCTION OF MEDIA:

6.7.1 After 14 days media shall be destructed as per sop or procedure.

6.8 FREQUENCY OF MEDIA FILLS (AMPULE/VIAL):

6.8.1 Initial Three consecutive Successful Run(all container size employed for filling on the machine) before start production for qualified area or equipment.

6.8.2 Then after once in 6+1 month of Single run any suitable volume for re-qualification of area or equipment.

6.8.3 If any change or modification of critical equipment or area then goes for single run media for any suitable volume before starting production.

6.8.4 If critical equipment or area is not in use for about one month then go for single run media for any suitable volume before starting production.

6.8.5 As and when required.

Also see: Media Fill Validation Test in Sterile Pharmaceuticals

Comments

Popular posts from this blog

Checklist for Audit in Sterile Area

Checklist for Audit in Sterile Area Checklist to verify the sterile manufacturing area before the audit. 1. Is there any SOP giving details for activity/ movement in the sterile area. 2. Verify the suitability of system for sanitization of hands. 3. Is there the SOP on entry-exit; gowning and de-gowning into the critical area available? 4. Verify the availability of following records:  Air pressure differential  Particulate monitoring  Temperature  Humidity  Microbiological monitoring by settle plate 5. Are these records available and maintained as per SOPs? 6. Check the above mentioned individual records. Are the entries within specifications mentioned in SOP/ MTPs. 7. Is the controlled copy of MTP pertaining to monitoring by settle plate available to plant person? 8. Are the location for settle plates identified and the chart/ layout available with plant? 9. Are the following areas monitored for parameters mentioned in step 4:  Air-lock to the critical ar...

MSDS (Material Safety Data Sheet)

MSDS   What Is an MSDS? An MSDS (Material Safety Data Sheet) is a document that provides detailed information about a chemical substance, including its hazards, safe handling procedures, storage requirements, and emergency measures. Today, MSDS is more commonly called SDS (Safety Data Sheet) under the Globally Harmonized System (GHS). Why MSDS / SDS Is Important Protect workers and employees Ensure safe handling and storage of chemicals Prevent accidents and injuries Provide emergency response information Comply with workplace safety regulations.  MSDS (Material Safety Data Sheet)  International safety standards.  GHS – Globally Harmonized System Full form: Globally Harmonized System of Classification and Labelling of Chemicals Develop by United Nations (UN).  OSHA (USA) – Hazard Communication Standard (HCS) Factories Act & Chemical Rules (India) Factories Act, 1948 Manufacture, Storage and Import of Hazardous Chemical Rules, 1989 REACH (Europe) REACH Regula...

Steps for HPLC Method Development

Steps for HPLC Method Development Know about the different steps of the HPLC analytical method development in pharmaceutical analysis. Analytical method development is considered as a critical process in pharmaceuticals. Availability of the different types of columns, operating  parameters, mobile phase composition, diluent and pH values make it critical to develop an analytical method. A good analytical method should be  simple, used column, mobile phase and buffer should be common. It can be done easily step by step. Following are the common HPLC method development steps. 1. Selection of HPLC Analytical Method 2. Selection of Chromatographic Conditions 3. Parameter Optimization 1. Selection of HPLC Analytical Method: First of all consult the literature that is available on the product. It will help you to understand the nature of  the product that will help to select the different parameters. A. Sample Preparation: Select method to prepare the sample according to its ...

Blow Fill Seal (BFS) and Form Fill Seal (FFS) Technology in Sterile Manufacturing

Blow Fill Seal (BFS) and Form Fill Seal (FFS) Technology in Sterile Manufacturing Know about sterile pharmaceutical production using Blow Fill Seal (BFS) and Form Fill Seal (FFS) technology. Both of these techniques are used to manufacture  sterile pharmaceutical products   as parenteral (LVP & SVP), infusions, ophthalmic and inhalation products. These are automated techniques to prepare sterile products. The basic concept of the  FFS   and BFS is to reduce the  contamination   by forming the container, filling and sealing in a closed sterile chamber of the machine. There is no personnel intervention to reduce the chances of the contamination during the manufacturing of sterile products. It gives more production at very low operational cost with the high assurance of  sterility . Blow fill seal technology is widely used and accepted by the various pharmaceutical regulatory authorities as US-FDA and MHRA. The system is being used for over 30 years a...