SOP for Calibration of HPLC (Low Pressure Quaternary Gradient Liquid Chromatograph)
Standard operating procedure of calibration of HPLC and its components like PUMP, Detector, Column Oven and injector.

1.0 OBJECTIVE
To lay down the procedure and frequency for calibration of HPLC.

2.0 SCOPE
This procedure is applicable to check the calibration of HPLC Quaternary Gradient system
(Model: LC 2010 C, LC 2010 C HT Make: Shimadzu) installed in Quality Control laboratory.

3.0 RESPONSIBILITY
3.1 Doing: Technical Assistant / Executive
3.2 Checking: Executive / Manager


4.0 ACCOUNTABILITY
Head of the department

5.0 FREQUENCY: ONCE IN A THREE MONTH.

6.0 PROCEDURE
6.1 Operate the instrument as per respective standard operating procedure.

6.2 For PUMP calibration
6.2.1 Set the flow rate to 1.0 ml/min.
6.2.2 Connect Resisto tubing 2 meter x 0.1 mm ID. in place of the column.
6.2.3 Start the pump and wait till equipment stabilized or a baseline is achieved.
6.2.4 Weigh the dry empty A grade measuring cylinder. Weigh & record the same.
6.2.5 Collect the water from the outlet of Resistor tubing for 5 minutes and weigh the measuring cylinder. Record the value.
6.2.6 Repeat the step 6.2.5 and continue the process for four repetitions.
6.2.7 Calculate the flow rate accuracy.

6.3 Wavelength accuracy of detector
6.3.1 Switch on the LC 2010 C, LC 2010 C HT by pressing the function key.
6.3.2 Press VP key.
6.3.3 Press Func Key and then Press F5.
6.3.4 Press 5 for the detector.
6.3.5 Press Right Arrow Key, to go to next screen (2/5).
6.3.6 Press Down Arrow Key seven times (make sure the cursor is blinking in ‘z wave’ parameter) press 1 enter.
6.3.7 Press Func and Press F5 for the exit.
6.3.8 Press Met Key.
6.3.9 Flush the flow cell with Methanol.
6.3.10 Open the port of Detector and pass the methanol with using the syringe for cleaning.
6.3.11 Press Zero Key and ensure that the absorbance is zero.
6.3.12 Inject 5 ml of 2 mg/100 ml solution of caffeine in methanol and changing the wavelength from 266 nm to 277 nm with the step of 1nm and
record the respective absorbance.
6.3.13 Record the wavelength that produces the maximum absorbance.
6.3.14 Record the results.

6.4 Reproducibility of Results
6.4.1 Prepare the mobile phase consisting of 70 % Methanol: 30 % Water.
6.4.2 Prepare a mixture of 0.1 % v/v Benzene and 0.1 % v/v Toluene in Methanol.
6.4.3 Set the flow rate of 1 ml/min.
6.4.4 Allow the system to be saturated with the mobile phase for at least 15 minutes before injecting the sample preparation.
6.4.5 Record the area and Relative Retention time of both benzene and toluene.
6.4.6 Calculate the RSD for both area and Relative Retention time respectively.
6.4.7 RSD of Area should not be more than 2.0 %. & RRT of Benzene & Toluene should not be more than 2.0 %.
6.4.8 Calculating the Tailing Factor ( Asymmetry), Resolution and Theoretical Plate and record.

6.5 Program files Check
6.5.1 Click start menu on the computer.
6.5.2 Go to Programs and select Chromatography.
6.5.3 Click Program Alteration Test.
6.5.4 Program files check window will open and Click Execute.
6.5.5 Software Checks all the files and gives the result.
6.5.6 If all the files are found good, Check is finished <<Integrity Check Passed>> will display.
6.5.7 Click notepad to see the status of files.
6.5.8 Click files in notepad window and select print for print out.
6.5.9 If some files are found not good, Check is finished <<Integrity Check Failed>> will display.
6.5.10 Click notepad to see the status of files.
6.5.11 In such case call service engineer.
6.5.12 Record the status of the program file.

6.6 Linearity test :
6.6.1 Prepare the mobile phase consisting of 70% Methanol: 30 % water.
6.6.2 Stock solution preparation: Prepare a mixture of 0.2 % Benzene and 0.2 % Toluene in methanol.
Make dilution in the following manner.
6.6.3 Take 5 ml, 10 ml, 15 ml, 20 ml of stock solution respectively in 25 ml of volumetric solution and makeup with methanol.
6.6.4 Set flow rate at 1.0 ml/min.
6.6.5 Allow the system to be saturated with mobile phase for at least 15 minutes before injecting the different solutions.
6.6.6 Record the area of both benzene and toluene.
6.6.7 Calculate the Linearity for both Benzene and Toluene respectively.

6.7 Carry Over test :
6.7.1 After completion of the Linearity test, inject 20μl of methanol.
6.7.2 Calculate the Carry Over test.

6.8 Quaternary gradient accuracy test :
6.8.1 Connect Resisto tubing ( 2 meter x 0.1 mm I.D.) to the instrument.
6.8.2 Prepare 0.3% Acetone in Purified Water.
6.8.3 Put mobile phase as follow into system.
A : Purified Water
B, C, D : 0.3 % Acetone in Purified Water.
6.8.4 Set flow rate to 1.0 ml/min.
6.8.5 Set wavelength to 254 nm.
6.8.6 Set time programme.
6.8.7 Adjust the baseline level to fit the full scale of the integrator.
6.8.8 Start the time programme for gradient performance.
6.8.9 Determine the signal level of 0% (B conc 0), 10 % (B conc 10), 50 % (B conc 50), 90% ( B conc 90), and 100 % (B conc 100).
6.8.10 Determine the signal level of 0% (C conc 0), 10 % (C conc 10), 50 % (C conc 50), 90% ( C conc 90), and 100 % (C conc 100).
6.8.11 Determine the signal level of 0% (D conc 0), 10 % (D conc 10), 50 % (D conc 50), 90% ( D conc 90), and 100 % (D conc 100)
6.8.12 Calculate concentration of B10, B50, B90, C10, C50, C90 and D10, D50, D90.
6.8.13 The performance of the instrument is satisfactory if the obtained RSD is within the given limit, otherwise follow the SOP.

6.9 Temperature Accuracy Test For Column Oven & Sample Cooler Tray:
6.9.1 Use a calibrated thermometers to measure the actual temperature in the oven & of sample cooler tray and compare it to set value
6.9.2 Procedure for Temperature Accuracy test for Column Oven.
6.9.2.1 Set the temperature accuracy is measured by using “ Performance Check ”
6.9.2.2 Press [VP] [F3 (VALID)] [2] [Enter] key, and select “ 5 oven Temperature” from the “ Performance Check ” menu.
6.9.2.3 Select the 40°C and 50°C temperature to be measured as instructed on the screen.
6.9.2.4 Use the clip to set a calibrated temperature sensor at the center of the lower step of heat block inside the device.
6.9.2.5 Press the [OVEN] key to start temperature adjustment. After the temperature has stabilized the temperature fluctuation amplitude will be
automatically measured for 5 minutes.
6.9.2.6 Record the Actual value against the calibrated thermometer value for different time intervals of 30, 45 and 60 minutes.
6.9.3 Procedure for Temperature Accuracy test for sample cooler tray
6.9.3.1 Press the [Seq], [F1] key, and enter “4” to “cooler temp” ( auto temperature setting parameter).
6.9.3.2 Enter 1 to “ SMPL.C” in other parameter section of Method screen.
6.9.3.3 Set the cooling rack ( R ) in the main unit & put the probe of calibrated thermometer inside the sample cooler tray.
6.9.3.4 Wait to stable at the set temperature since it normally takes around 90 min.
6.9.3.5 Press [VP] F3 (VALID)] [2] [Enter] key, and select “ 11. Cooler Temperature” from the “ Performance Check”.
6.9.3.6 Record the actual value against the calibrated thermometer value for different time intervals of 30, 45 and 60 minutes.

7.0 ABBREVIATION
RRT: Relative Retention Time
%: Percentage.
ml: milliliter
RSD: Relative Standard Deviation.
RT: Retention Time
v/v: Volume / Volume
Also see: Relative Response Factor (RRF) Calculation in HPLC

Comments

Popular posts from this blog

Checklist for Audit in Sterile Area

Checklist for Audit in Sterile Area Checklist to verify the sterile manufacturing area before the audit. 1. Is there any SOP giving details for activity/ movement in the sterile area. 2. Verify the suitability of system for sanitization of hands. 3. Is there the SOP on entry-exit; gowning and de-gowning into the critical area available? 4. Verify the availability of following records:  Air pressure differential  Particulate monitoring  Temperature  Humidity  Microbiological monitoring by settle plate 5. Are these records available and maintained as per SOPs? 6. Check the above mentioned individual records. Are the entries within specifications mentioned in SOP/ MTPs. 7. Is the controlled copy of MTP pertaining to monitoring by settle plate available to plant person? 8. Are the location for settle plates identified and the chart/ layout available with plant? 9. Are the following areas monitored for parameters mentioned in step 4:  Air-lock to the critical ar...

Aseptic Filling Process (Media Fill) Validation Protocol in Sterile Pharmaceuticals

Aseptic Filling Process (Media Fill) Validation Protocol in Sterile Pharmaceuticals Validation of sterile manufacturing process by media fill validation test as per PICS guidelines for aseptic validation or aseptic process simulation. 1.0 OBJECTIVE 1.1 To define procedures for validating and maintaining the validation of all aseptic filling processes and qualification of the quality of the product by system/facility/equipment. 2.0 SCOPE 2.1 This procedure applies to all aseptically filled sterile products intended for human use. 3.0 RESPONSIBILITY 3.1 The Validation personnel coordinate the aseptic filling validation program and write the sterile media aseptic filling processes reports. 3.2 Manufacturing personnel the sterile media aseptic filling processes and performs the sterile media fills. 3.3 QA personnel perform the sampling and assist with the IPQA monitoring required for each sterile media fill. 3.4 QC personnel to perform the testing and assist with the monitoring required fo...

Blow Fill Seal (BFS) and Form Fill Seal (FFS) Technology in Sterile Manufacturing

Blow Fill Seal (BFS) and Form Fill Seal (FFS) Technology in Sterile Manufacturing Know about sterile pharmaceutical production using Blow Fill Seal (BFS) and Form Fill Seal (FFS) technology. Both of these techniques are used to manufacture  sterile pharmaceutical products   as parenteral (LVP & SVP), infusions, ophthalmic and inhalation products. These are automated techniques to prepare sterile products. The basic concept of the  FFS   and BFS is to reduce the  contamination   by forming the container, filling and sealing in a closed sterile chamber of the machine. There is no personnel intervention to reduce the chances of the contamination during the manufacturing of sterile products. It gives more production at very low operational cost with the high assurance of  sterility . Blow fill seal technology is widely used and accepted by the various pharmaceutical regulatory authorities as US-FDA and MHRA. The system is being used for over 30 years a...