Skip to main content

HPLC Calibration Procedure

HPLC Calibration Procedure :
Know the procedure to calibrate the High Performance Liquid Chromatography ( HPLC) including leakage test, flow rate, reproducibility and linearity, lamp energy and pump pressure drop in Pharmaceutical Quality Control.

Check HPLC chromatography (Pump) for the following:


  1. Checkpoint: Leakage test (By Pressure Drop).
  2. Flow rate calibration

CHECKPOINT: LEAKAGE TEST (BY PRESSURE DROP):
1. Ensure that, the instrument is ready for calibration and Start-up procedure is followed.
2. Place inlet tubing of the Pump into the Water HPLC grade through a suction filter.
3. Allow the mobile phase to flow for about 5 min.
4. Block Pump outlet with the block screw.

5. The pressure rises and on crossing the 300 bar, “ERROR P-MAX” appears on the display window. Note the time. Press “CE” key and observe the pressure drop for 5 min.
6. After 5 min., record the pressure in the Calibration Log.
7. Make entry of the HPLC column usage in the Column Usage Log Register.
8. Make entry of the user into the Instrument Usage Log Register.
9. Compare the result for its compliance with the limit given in the Calibration Log and put the remark regarding the HPLC chromatography Calibration Status.
10. In case of non-compliance, follow the Maintenance Program.

FLOW RATE CALIBRATION
1. Ensure that, the instrument is ready for calibration and Start-up procedure is followed.
2. Ensure that, the Pump is passing the “Leakage Test (By Pressure Drop)”.
3. Keep the Drain tube in such a way that the mobile phase (Water) drops falls into 10 ml clean, dry volumetric flask without touching the walls of the flask and start immediately the stopwatch when the first drop falls into the flask.
4. Wait till the collected mobile phase reaches 10 ml mark of the volumetric flask and Stop the stopwatch.
5. Record the time required to collect the 10 ml mobile phase in the calibration log.
6. Repeat the procedure for 1.0 ml, 1.5 ml and 2.0 ml/ min. flow rates.
7. Repeat step 3 to 6 but using methanol HPLC grade as mobile phase instead of water.
8. Compare the results for its compliance with limits given in the Calibration Log and put the remark regarding the Calibration Status.
9. Make entry of the usage of the instrument and HPLC column in the Instrument Usage Log Register and Column Usage Log Register respectively.
10. Prepare Calibration Status Label and display on the instrument at the designated place.
11. In case of non-compliance, follow the Maintenance Program

REPRODUCIBILITY AND LINEARITY OF INJECTION VOLUME
Solution Mixer Preparation
1. Take clean and dry 50 ml volumetric flask.
2. Pipette out 1.0 ml of Benzene and Toluene into the clean and dry 50 ml volumetric flask.
3. Make up the volume to 50 ml with Methanol and mix well

Chromatographic Condition
HPLC Column: ODS C18, (25 cm x 4.6 mm ID, 5 μm)
Mobile Phase: Methanol : Water (70 : 30)
Flow Rate: 1.0 ml/min.
Wavelength: 254 nm
Injection Volume: 20 μl

Calibration :
1. Ensure that, the instrument is ready for calibration and Start-up procedure is followed.
2. Ensure that, the instrument is set according to the Chromatographic conditions.
3. Follow the Instrument Operating procedure, Inject 10 μl in triplicate and record the chromatograms.
4. Repeat the injection of the above solution by injecting 15, 20, 25, 30 μl in triplicate.
5. Take the print out of the chromatograms and attach to the Calibration Log.
6. Record the Area and Retention times of the of the Benzene and Toluene peaks in the Calibration Log.
7. Make entry of the usage of the instrument and HPLC column in the Instrument Usage Log Register and Column Usage Log Register respectively.
8. Plot the curve for the area corresponding to Benzene to Toluene peaks v/s injection volume, Find out the RSD (reproducibility) and record in the Calibration Log.
9. Find out the Correlation coefficient “r2” for each peak at five levels and record in the Calibration Log.
10. Compare the result for its compliance with the limit given in the Calibration Log and put the remark regarding the Calibration Status.
11. Prepare Calibration Status Label and display on the instrument at the designated place.
12. In case of non-compliance, follow the Maintenance Program.

CHECKPOINT: D2 LAMP ENERGY CHECK (DETECTOR)
1. Ensure that, the instrument is ready for calibration and Start-up procedure is followed.
2. On the Detector’s display window, some values at the functions “l(nm)”, “abs(AU)”, “range(AUFS)” and “lamp” appears.
3. On the display, previously set value blinks at “l(nm)” function, enter the wavelength to 254 nm by pressing numeric keys.
4. Press “Func Back” key and select “lamp” functions, enter 1 to select D2 lamp.
5. Further press “Func Back” key till “REF EN ” appears.
6. Record the Reference Energy of the D2 Lamp at 254 nm in the Calibration Log.
7 Compare the result for its compliance with the limit given in the Calibration Log and put the remark regarding the Calibration Status.
8. In case of non-compliance, follow the Maintenance Program.

LINEARITY OF DETECTOR RESPONSE :
Solution Mixer Preparation:
1. Take three clean and dry 50 ml volumetric flasks
2. Pipette out ml of Benzene and Toluene as per the following table into the clean, dry 50 ml volumetric flask separately.
3. Make up the volume to 50 ml with Methanol and mix well.

Solution (Level)
ml of Benzene to be taken
Ml of Toluene to be taken
To be Diluted with Methanol to
1
0.5
1.0
50 ml
2
1.0
1.0
50 ml
3
1.5
1.0
50 ml

Chromatographic Condition
1.    Set each of the equipment to the following parameters as per the Equipment’s Parameter Setup Procedure.
HPLC Column: ODS C18, (25 cm x 4.6 mm ID, 5 m)
Mobile Phase: Methanol : Water (70 : 30)
Flow Rate: 1.0 ml/min.
Wavelength: 254 nm
Injection Volume: 20 μl

Calibration
1. Ensure that, the instrument is ready for calibration and Start-up procedure is followed.
2. Ensure that, the instrument is set according to the Chromatographic conditions.
3. Follow the Instrument Operating procedure, Inject each of the Solutions in triplicate and record the chromatograms.
4. Take the print out of the chromatograms and attach to the Calibration Log.
5. Record the Area and Retention times of the of the Benzene and Toluene peaks in the Calibration Log.
6. Make entry of the usage of the instrument and HPLC column in the Instrument Usage Log Register and Column Usage Log Register respectively.
7. Calculate the Area Ratio of Benzene to Toluene, Find out the Mean value of the ratios and record in the Calibration Log.
8. Find out the Correlation coefficient “r2” from the Mean area ratio values of the three levels.and record in the Calibration Log.
8. Plot the curve for area ratio corresponding to Benzene to Toluene peaks v/s concentration, Find out the RSD (reproducibility) and record in theCalibration Log.
9. Compare the result for its compliance against limit given in the Calibration Log and put the remark regarding the Calibration Status.
10. Prepare Calibration Status Label and display on the instrument at the designated place.
11. In case of non-compliance, follow the Maintenance Program.

CALIBRATION OF HPLC ( Liquid Chromatography ) PUMP

Model No. :                                                               Make :
Instrument Code No.

a.    Checkpoint : Leakage Check Test (Pressure Drop)
P. Max to be set
P. Max Set
Start Time
Stop Time
Total Time
Pressure Observed
Remark
300 bar






Flow Rate Calibration :
Flow Rate (ml / min.)
Time required to collect 10 ml mobile phase
Limit
Remarks
Theoretical
(in sec.)
Actual with water
Actual with Methanol
0.5 ml
1200


1194 – 1206

1.0 ml
600


594 – 606

1.5 ml
450


443 – 457

2.0 ml
300


294 - 306

Calibration Status : Satisfactory / Not Satisfactory

Next Calibration Due :
Calibrated By :                                                               Checked By :
Date :                                                                              Date :

CALIBRATION OF INJECTOR
Model:                                                                                       Make:
Instrument Code No.:

a.     Preparation of solvent mixture :
Taken _____ ml (1.0 ml) of the ________ (Benzene) and ________ (Toluene) in to a _____ ml (50 ml) clean, dry volumetric flask, make up the volume with methanol, mixed well.


b.    Chromatographic Condition :
Parameters
Test Condition
Applied Condition
Mobile Phase
Methanol: Water (70: 30)

HPLC Column
ODS C18, (25 cm x 4.6 mm ID, 5 μm)

Flow Rate
1.0 ml / min.

Wavelength
254 nm


Injection volume
Injection -1
Injection -2
Injection -3
Mean
%RSD:(Not more than 2.0%)
Benzene
Toluene
Benzene
Toluene
Benzene
Toluene


10 ml
RT








Area








15 ml
RT








Area








20 ml
RT








Area








25 ml
RT








Area








30 ml
RT








Area








Coefficient of co-relations : r2 : _____________ (Limit : NLT : 0.999)

Calibration Status : Satisfactory / Not Satisfactory
Next Calibration Due :
Calibrated By:                                                                             Checked By :
Date:                                                                                           Date :

CALIBRATION OF DETECTOR
 Model :                                                                                        Make :
Instrument Code No. :
D2 LAMP ENERGY CHECK :
Wavelength to be set
Wavelength Set
                Reference Energy Observed
Limit
254 nm


Not less than 200
LINEARITY OF DETECTOR RESPONSE :
a. Solution Preparation :
Solution
ml of Benzene
ml of Toluene
in Methanol
(Level)
To be taken
Taken
To be taken
Taken
To be diluted
Diluted to
1
0.5 ml

1.0 ml

50 ml

2
1.0 ml

1.0 ml

50 ml

3
1.5 ml

1.0 ml

50 ml


Mobile Phase Preparation: Taken ______ ml of ___________ (Methanol HPLC grade) in to a _______ ml clean, dry volumetric flask, added ____________ ml of___________(Water HPLC grade), mixed well, allowed to cool to room temperature. Filtered through 0.45/0.22 m membrane filter,  degassed  for _____ min. by vacuum  / sonication.

Parameters
Test Condition
Applied Condition
Mobile Phase
Methanol: Water ( 70: 30)

HPLC Column
ODS C18, (25 cm x 4.6 mm ID, 5 μm)

Flow Rate
1.0 ml / min.

Wavelength
254 nm

Range
1.0 AUFS

Injection Volume
20 μl


Solution(Level)
Benzene Area
Toluene Area
Area Ratio(Benzene/Toluene)
RT
Area
RT
Area
1

















Mean :





% RSD :





2

















Mean :





% RSD :





3

















Mean :





% RSD :






Linearity of Ratio : Correlation Co-efficient r2 :
Limit : % RSD of the Retention Time : Not more than 2.0%
Linearity of Ratio : Correlation Co-efficient r2 : Not less than 0.999

Calibration Status : Satisfactory / Not Satisfactory
Next Calibration Due :
Calibrated By :                                                                             Checked By :
Date :                                                                                            Date :


Comments

Popular posts from this blog

Checklist for Audit in Sterile Area

Checklist for Audit in Sterile Area Checklist to verify the sterile manufacturing area before the audit. 1. Is there any SOP giving details for activity/ movement in the sterile area. 2. Verify the suitability of system for sanitization of hands. 3. Is there the SOP on entry-exit; gowning and de-gowning into the critical area available? 4. Verify the availability of following records:  Air pressure differential  Particulate monitoring  Temperature  Humidity  Microbiological monitoring by settle plate 5. Are these records available and maintained as per SOPs? 6. Check the above mentioned individual records. Are the entries within specifications mentioned in SOP/ MTPs. 7. Is the controlled copy of MTP pertaining to monitoring by settle plate available to plant person? 8. Are the location for settle plates identified and the chart/ layout available with plant? 9. Are the following areas monitored for parameters mentioned in step 4:  Air-lock to the critical ar...

Aseptic Filling Process (Media Fill) Validation Protocol in Sterile Pharmaceuticals

Aseptic Filling Process (Media Fill) Validation Protocol in Sterile Pharmaceuticals Validation of sterile manufacturing process by media fill validation test as per PICS guidelines for aseptic validation or aseptic process simulation. 1.0 OBJECTIVE 1.1 To define procedures for validating and maintaining the validation of all aseptic filling processes and qualification of the quality of the product by system/facility/equipment. 2.0 SCOPE 2.1 This procedure applies to all aseptically filled sterile products intended for human use. 3.0 RESPONSIBILITY 3.1 The Validation personnel coordinate the aseptic filling validation program and write the sterile media aseptic filling processes reports. 3.2 Manufacturing personnel the sterile media aseptic filling processes and performs the sterile media fills. 3.3 QA personnel perform the sampling and assist with the IPQA monitoring required for each sterile media fill. 3.4 QC personnel to perform the testing and assist with the monitoring required fo...

MSDS (Material Safety Data Sheet)

MSDS   What Is an MSDS? An MSDS (Material Safety Data Sheet) is a document that provides detailed information about a chemical substance, including its hazards, safe handling procedures, storage requirements, and emergency measures. Today, MSDS is more commonly called SDS (Safety Data Sheet) under the Globally Harmonized System (GHS). Why MSDS / SDS Is Important Protect workers and employees Ensure safe handling and storage of chemicals Prevent accidents and injuries Provide emergency response information Comply with workplace safety regulations.  MSDS (Material Safety Data Sheet)  International safety standards.  GHS – Globally Harmonized System Full form: Globally Harmonized System of Classification and Labelling of Chemicals Develop by United Nations (UN).  OSHA (USA) – Hazard Communication Standard (HCS) Factories Act & Chemical Rules (India) Factories Act, 1948 Manufacture, Storage and Import of Hazardous Chemical Rules, 1989 REACH (Europe) REACH Regula...

Steps for HPLC Method Development

Steps for HPLC Method Development Know about the different steps of the HPLC analytical method development in pharmaceutical analysis. Analytical method development is considered as a critical process in pharmaceuticals. Availability of the different types of columns, operating  parameters, mobile phase composition, diluent and pH values make it critical to develop an analytical method. A good analytical method should be  simple, used column, mobile phase and buffer should be common. It can be done easily step by step. Following are the common HPLC method development steps. 1. Selection of HPLC Analytical Method 2. Selection of Chromatographic Conditions 3. Parameter Optimization 1. Selection of HPLC Analytical Method: First of all consult the literature that is available on the product. It will help you to understand the nature of  the product that will help to select the different parameters. A. Sample Preparation: Select method to prepare the sample according to its ...

Blow Fill Seal (BFS) and Form Fill Seal (FFS) Technology in Sterile Manufacturing

Blow Fill Seal (BFS) and Form Fill Seal (FFS) Technology in Sterile Manufacturing Know about sterile pharmaceutical production using Blow Fill Seal (BFS) and Form Fill Seal (FFS) technology. Both of these techniques are used to manufacture  sterile pharmaceutical products   as parenteral (LVP & SVP), infusions, ophthalmic and inhalation products. These are automated techniques to prepare sterile products. The basic concept of the  FFS   and BFS is to reduce the  contamination   by forming the container, filling and sealing in a closed sterile chamber of the machine. There is no personnel intervention to reduce the chances of the contamination during the manufacturing of sterile products. It gives more production at very low operational cost with the high assurance of  sterility . Blow fill seal technology is widely used and accepted by the various pharmaceutical regulatory authorities as US-FDA and MHRA. The system is being used for over 30 years a...