HPLC System Software Validation Plan

HPLC System Software Validation Plan
HPLC software are required to be validated to assure the performance of the system. The same should be validated before its use in analysis.

Introduction

Softwares are used to operate and interpret the analysis results in HPLC systems. There are many softwares those are used in different HPLC systems. All softwares are required to validate to ensure the accuracy of the produced data.

Software Validation is a requirement of GxP system. This software validation is intended to provide details of activities carried out and results therein to support the qualification and operation of Computer System, which are                                                                         connected with the HPLC systems.

Purpose
This validation plan is to outline the activities carried out for software validation of HPLC system as per the approach described in the validation plan.

Scope
This software validation is limited to HPLC system its unique components and control system that defines the hardware and software.

Policy Compliance
This document is being written to comply with regulatory requirements for software validation of HPLC systems.

Validation is a critical tool to assure the quality of computer system performance. Computer system software validation increases the reliability of systems, resulting in fewer errors and less risk to process and data integrity. Computer system validation also reduces long-term system and project
costs by minimizing the cost of maintenance and rework.

Approach
This validation is conducted as a retrospective validation. The validation plan documents the process that is to demonstrate the equipment performs as intended.

Based on the user requirement, design, test plan and test scripts are developed and verified against the existing system and software validation package carried out at the site by the supplier.

Roles & Responsibilities
Specific responsibilities related to the software validation of the HPLC systems outlined in the Validation Plan. In general, the activities associated with this project, are the responsibility of the individuals and groups listed in the Validation Plan.

Regulatory Status
This system is GxP system. System components that purchased are identified in the Qualification carried out by the supplier in the presence of system custodian.

Deliverables
The following is the list of deliverable developed to comply with the policy standards.

User Requirements
The user requirements developed. User requirements defined for the existing system based on the critical functions performed by the system.

Vendor Management
The IQ/OQ/PQ has been outsourced to the manufacturer. The documents are verified and found satisfactory. Qualification of HPLC systems for hardware and software at the site is carried out in the presence of system custodian. The detailed training of hardware and software provided to users. Troubleshooting and maintenance services are provided by the supplier as and when required. The manufacturer is a primary supplier of the HPLC system and provides technical support and services as well as technical assistance. The manufacturer had provided us a certificate of compliance.

Design
A design document developed with the help of the available reference manuals, literature. We have collected the information from the documentation provided by the manufacturer of the respective system, but we have also used the current implementation of the system to understand how the system was built.

Application Software
This system is custom off the shelf, no macro has been developed. Therefore no source code review is performed. The manufacturer has a coding standard, which is unique, traceable and secure. It also has good development practices for application software manufacturing.

Testing
We have developed a test plan consisting test protocol with the help of the current operating procedure documents and testing performed accordingly.

Security Plan
HPLC instrument located in Quality Control Instrument room having an adequate facility (e.g. proper air-conditioned and UPS power supply) at QC Department, SPI, Dadra. QC Department is having a duplicate lock & key facility. One key is kept with the Security Department and another with QC In-charge. Only In-charge is authorized to open the QC room. Operating of HPLC systems are allowed only to authorized personnel.

Backup and restore
We have developed a Backup and restore procedure with the help of the current backup system.

Training
The existing people have been trained for use of Chromatographic Systems and new joinee shall be trained for the operation of HPLC systems.
Training record shall manage as per the training guideline of Quality Control.

Change Control
Change control procedure is followed as per SOP.

Periodic Review
The periodic review of documentation and status of validation shall be conducted. The evidence of the review shall be documented separately.

Comments

Popular posts from this blog

Checklist for Audit in Sterile Area

Checklist for Audit in Sterile Area Checklist to verify the sterile manufacturing area before the audit. 1. Is there any SOP giving details for activity/ movement in the sterile area. 2. Verify the suitability of system for sanitization of hands. 3. Is there the SOP on entry-exit; gowning and de-gowning into the critical area available? 4. Verify the availability of following records:  Air pressure differential  Particulate monitoring  Temperature  Humidity  Microbiological monitoring by settle plate 5. Are these records available and maintained as per SOPs? 6. Check the above mentioned individual records. Are the entries within specifications mentioned in SOP/ MTPs. 7. Is the controlled copy of MTP pertaining to monitoring by settle plate available to plant person? 8. Are the location for settle plates identified and the chart/ layout available with plant? 9. Are the following areas monitored for parameters mentioned in step 4:  Air-lock to the critical ar...

Aseptic Filling Process (Media Fill) Validation Protocol in Sterile Pharmaceuticals

Aseptic Filling Process (Media Fill) Validation Protocol in Sterile Pharmaceuticals Validation of sterile manufacturing process by media fill validation test as per PICS guidelines for aseptic validation or aseptic process simulation. 1.0 OBJECTIVE 1.1 To define procedures for validating and maintaining the validation of all aseptic filling processes and qualification of the quality of the product by system/facility/equipment. 2.0 SCOPE 2.1 This procedure applies to all aseptically filled sterile products intended for human use. 3.0 RESPONSIBILITY 3.1 The Validation personnel coordinate the aseptic filling validation program and write the sterile media aseptic filling processes reports. 3.2 Manufacturing personnel the sterile media aseptic filling processes and performs the sterile media fills. 3.3 QA personnel perform the sampling and assist with the IPQA monitoring required for each sterile media fill. 3.4 QC personnel to perform the testing and assist with the monitoring required fo...

MSDS (Material Safety Data Sheet)

MSDS   What Is an MSDS? An MSDS (Material Safety Data Sheet) is a document that provides detailed information about a chemical substance, including its hazards, safe handling procedures, storage requirements, and emergency measures. Today, MSDS is more commonly called SDS (Safety Data Sheet) under the Globally Harmonized System (GHS). Why MSDS / SDS Is Important Protect workers and employees Ensure safe handling and storage of chemicals Prevent accidents and injuries Provide emergency response information Comply with workplace safety regulations.  MSDS (Material Safety Data Sheet)  International safety standards.  GHS – Globally Harmonized System Full form: Globally Harmonized System of Classification and Labelling of Chemicals Develop by United Nations (UN).  OSHA (USA) – Hazard Communication Standard (HCS) Factories Act & Chemical Rules (India) Factories Act, 1948 Manufacture, Storage and Import of Hazardous Chemical Rules, 1989 REACH (Europe) REACH Regula...

Steps for HPLC Method Development

Steps for HPLC Method Development Know about the different steps of the HPLC analytical method development in pharmaceutical analysis. Analytical method development is considered as a critical process in pharmaceuticals. Availability of the different types of columns, operating  parameters, mobile phase composition, diluent and pH values make it critical to develop an analytical method. A good analytical method should be  simple, used column, mobile phase and buffer should be common. It can be done easily step by step. Following are the common HPLC method development steps. 1. Selection of HPLC Analytical Method 2. Selection of Chromatographic Conditions 3. Parameter Optimization 1. Selection of HPLC Analytical Method: First of all consult the literature that is available on the product. It will help you to understand the nature of  the product that will help to select the different parameters. A. Sample Preparation: Select method to prepare the sample according to its ...

Blow Fill Seal (BFS) and Form Fill Seal (FFS) Technology in Sterile Manufacturing

Blow Fill Seal (BFS) and Form Fill Seal (FFS) Technology in Sterile Manufacturing Know about sterile pharmaceutical production using Blow Fill Seal (BFS) and Form Fill Seal (FFS) technology. Both of these techniques are used to manufacture  sterile pharmaceutical products   as parenteral (LVP & SVP), infusions, ophthalmic and inhalation products. These are automated techniques to prepare sterile products. The basic concept of the  FFS   and BFS is to reduce the  contamination   by forming the container, filling and sealing in a closed sterile chamber of the machine. There is no personnel intervention to reduce the chances of the contamination during the manufacturing of sterile products. It gives more production at very low operational cost with the high assurance of  sterility . Blow fill seal technology is widely used and accepted by the various pharmaceutical regulatory authorities as US-FDA and MHRA. The system is being used for over 30 years a...