Skip to main content

SOP for Sampling and Testing Schedule of Purified Water

 SOP for Sampling and Testing Schedule of Purified Water

Standard operating procedure for schedule & frequency for testing, locations for sampling and type of analysis required to be carried out for the Purified water for monitoring & assurance of its quality.

1.0 OBJECTIVE

To define the schedule & frequency for testing, locations for sampling and type of analysis required to be carried out for the Purified                                                                                      water for monitoring & assurance of its quality.

2.0 SCOPE

This Procedure is applicable during sampling and analysis of purified water.

3.0 RESPONSIBILITY

3.1 Doing: Tech.Assistant (Microbiologist)

3.2 Checking: Executive/Manager

4.0 ACCOUNTABILITY

Head of the Department

5.0 PROCEDURE

Remark: Ensure visually that container for sample collection is washed and dried properly.

5.1 Collection of a Purified water sample for chemical analysis.

5.1.1 First collect the sample for chemical analysis.

5.1.2 Flush water from the tap for 30 seconds.

5.1.3 Rinse the sampling bottles three times with actual sample before sampling for chemical analysis.

5.1.4 Collect approx. 500 ml sample of Purified water in b the ttle

Note: Sample should be filled up to the neck of sampling bottle for chemical analysis.

This is to ensure that air in the empty portion of the sampling bottle should not get dissolved in the water sample.

5.2 Collection of Purified water sample for Total Organic Carbon (TOC) test.

5.2.1 Use airtight container for collection of Purified water sample for TOC analysis.

5.2.2 Wash the container prior to sampling with 2N HCl for 2-3 times.

5.2.3 Follow pt. 5.1.2 & then 5.1.3.

5.2.4 Collect about 60-70 ml of sample & close the sample container tightly.

Note: (1) Take precaution to prevent atmospheric carbon dioxide being dissolved in sample.

(2) In case if TOC analyser is out of order, the sample can be release on base of Oxidisable substance test Result.

5.3 Collect the samples of purified water and test accordingly as per Annexure.

5.3.1 Use suitable container for collection of Purified water sample. This to be sterilized as under.

→ Fill the sample collection container with PW & keep the cap tightly.

→ Wrap it with wrapping/butter paper

→ Sterilise it at 121°C for 30min.

5.3.2 Discard the PW filled in sterilized sample container just before collection of purified water sample for analysis.

5.3.3 Collect about 15 ml of purified water sample in sterilized container and close tightly.

5.4 Collection of purified water sample for Bacterial Endotoxin Test (BET)

5.4.1 Use Depyrogenated glass container for collection of purified water sample.

5.4.2 Collect about 15ml purified water sample for BET.

Remark: (1) Send the sample within 1hr. to microbiology Lab for BET & Microbiological test.

(2) After sampling it is recommended that, the water sample is keep at 2-8 °C.

If analysis cannot be performed within 8 hours after sampling, the sample must be kept at 2-8°C.

The microbiological testing of a sample must be performed no later than 24 hours after sampling and within the time limit test should be started.

5.5 Purified water frequency of testing

Location                 Test to be carried out                                      Frequency

UF Outlet                 Chemical test, pH, Conductivity, TVC, TOC.     Daily

User End Point        Total viable count, pH, Conductivity, TOC         Once in a week

Note: Complete analysis: As per specification 

6.0 ABBREVIATIONS

°C = Degree centigrade

TVC = Total viable count

TOC = Total organic carbon

Q.C = Quality control department

Comments

Popular posts from this blog

Checklist for Audit in Sterile Area

Checklist for Audit in Sterile Area Checklist to verify the sterile manufacturing area before the audit. 1. Is there any SOP giving details for activity/ movement in the sterile area. 2. Verify the suitability of system for sanitization of hands. 3. Is there the SOP on entry-exit; gowning and de-gowning into the critical area available? 4. Verify the availability of following records:  Air pressure differential  Particulate monitoring  Temperature  Humidity  Microbiological monitoring by settle plate 5. Are these records available and maintained as per SOPs? 6. Check the above mentioned individual records. Are the entries within specifications mentioned in SOP/ MTPs. 7. Is the controlled copy of MTP pertaining to monitoring by settle plate available to plant person? 8. Are the location for settle plates identified and the chart/ layout available with plant? 9. Are the following areas monitored for parameters mentioned in step 4:  Air-lock to the critical ar...

Aseptic Filling Process (Media Fill) Validation Protocol in Sterile Pharmaceuticals

Aseptic Filling Process (Media Fill) Validation Protocol in Sterile Pharmaceuticals Validation of sterile manufacturing process by media fill validation test as per PICS guidelines for aseptic validation or aseptic process simulation. 1.0 OBJECTIVE 1.1 To define procedures for validating and maintaining the validation of all aseptic filling processes and qualification of the quality of the product by system/facility/equipment. 2.0 SCOPE 2.1 This procedure applies to all aseptically filled sterile products intended for human use. 3.0 RESPONSIBILITY 3.1 The Validation personnel coordinate the aseptic filling validation program and write the sterile media aseptic filling processes reports. 3.2 Manufacturing personnel the sterile media aseptic filling processes and performs the sterile media fills. 3.3 QA personnel perform the sampling and assist with the IPQA monitoring required for each sterile media fill. 3.4 QC personnel to perform the testing and assist with the monitoring required fo...

MSDS (Material Safety Data Sheet)

MSDS   What Is an MSDS? An MSDS (Material Safety Data Sheet) is a document that provides detailed information about a chemical substance, including its hazards, safe handling procedures, storage requirements, and emergency measures. Today, MSDS is more commonly called SDS (Safety Data Sheet) under the Globally Harmonized System (GHS). Why MSDS / SDS Is Important Protect workers and employees Ensure safe handling and storage of chemicals Prevent accidents and injuries Provide emergency response information Comply with workplace safety regulations.  MSDS (Material Safety Data Sheet)  International safety standards.  GHS – Globally Harmonized System Full form: Globally Harmonized System of Classification and Labelling of Chemicals Develop by United Nations (UN).  OSHA (USA) – Hazard Communication Standard (HCS) Factories Act & Chemical Rules (India) Factories Act, 1948 Manufacture, Storage and Import of Hazardous Chemical Rules, 1989 REACH (Europe) REACH Regula...

Steps for HPLC Method Development

Steps for HPLC Method Development Know about the different steps of the HPLC analytical method development in pharmaceutical analysis. Analytical method development is considered as a critical process in pharmaceuticals. Availability of the different types of columns, operating  parameters, mobile phase composition, diluent and pH values make it critical to develop an analytical method. A good analytical method should be  simple, used column, mobile phase and buffer should be common. It can be done easily step by step. Following are the common HPLC method development steps. 1. Selection of HPLC Analytical Method 2. Selection of Chromatographic Conditions 3. Parameter Optimization 1. Selection of HPLC Analytical Method: First of all consult the literature that is available on the product. It will help you to understand the nature of  the product that will help to select the different parameters. A. Sample Preparation: Select method to prepare the sample according to its ...

Blow Fill Seal (BFS) and Form Fill Seal (FFS) Technology in Sterile Manufacturing

Blow Fill Seal (BFS) and Form Fill Seal (FFS) Technology in Sterile Manufacturing Know about sterile pharmaceutical production using Blow Fill Seal (BFS) and Form Fill Seal (FFS) technology. Both of these techniques are used to manufacture  sterile pharmaceutical products   as parenteral (LVP & SVP), infusions, ophthalmic and inhalation products. These are automated techniques to prepare sterile products. The basic concept of the  FFS   and BFS is to reduce the  contamination   by forming the container, filling and sealing in a closed sterile chamber of the machine. There is no personnel intervention to reduce the chances of the contamination during the manufacturing of sterile products. It gives more production at very low operational cost with the high assurance of  sterility . Blow fill seal technology is widely used and accepted by the various pharmaceutical regulatory authorities as US-FDA and MHRA. The system is being used for over 30 years a...