Skip to main content

Sterility Testing Procedure

Sterility Testing Procedure

Learn how to perform the test for sterility for sterile pharmaceutical products by Membrane Filtration Method.

1.0 Equipment Required

Sterility Testing Procedure, what is sterility test
Sterility Testing Procedure

¤ LAF

¤ Manifold holder assembly

¤ Vacuum pump

¤ Forceps

¤ Scissor



2.0 Material Required

© Sample

© Sterile 0.45µ Membrane filter

© Sterile FTM

© Sterile SCDM

© Sterile 0.1% w/v Peptone water

© 70% sterile IPA Solution

© Gas burner

© Sterile Filtration assembly

© Sterile SCDA plates

© Sterile Swab

3.0 Utilities Required

Vacuum Pump

4.0 Procedure:

Membrane Filtration Method

→ Sample for Finished Products: Collect the samples to be tested for sterility as per SOP. Out of the whole sample, randomly select 20 articles (as per Pharmacopoeia) of each lot of batch for both LVP and SVP terminally sterilized products and for aseptically filled products.

→ Sample for Intermediates: Randomly collect 16 pre-sterilized bottle samples to be tested for sterility from LVP bottle pack machine (When change over the new Product collect the 16 bottle samples of 1st batch ) separately in such a way that each bottle should represent each cavity of the mold of bottle pack machine.

→ Transfer the samples to QC department in clean plastic crates.

→ Wipe the sample article individually with 70% IPA solution and keep in a clean S.S tray marked with Product Name, Batch No. and Lot No, and then transfer the samples to the sterility room through clean pass box for performing sterility.

→ Prepare the media tubes (FTM and SCDM) as per the SOP for preparation of culture media, dispense 100-100 ml quantity in test tubes & plug them. Sterilize both the media at 121ºC and 15 psi pressure for 20 minutes as per SOP for Media Sterilization by Autoclaving.

→ After autoclaving Label the tubes with Name of Media, Media Batch No. and pre-incubate the media tubes at appropriate temperature i.e. SCDM tubes at 20 to 25ºC whereas FTGM tubes at 30 to 35ºC for 24 - 48 hrs before subjecting them for sterility operations.

→ Autoclave Dress, S.S. cups, Receptacle unit, Scissors and forceps in a stainless steel container at 121ºC temperature and 15psi pressure for 30 minutes as per SOP.

→ After Sterilization cool the contents and aseptically transfer in an S.S. container to cleaned Pass box.

→ Transfer the pre-incubated sterile media tubes, SCDA plates, sterile swabs, sterilized manifold, and 0.1% w/v peptone water to the sterility test room through pass box.

→ Enter in sterility room as per the S OP for Entry / Exit procedure for Sterility Room.

→ Start the LAF as per SOP for operating instruction for LAF.

→ Wipe out all samples to be tested for sterility with 70% IPA solution.

→ Before starting sterility test, expose the SCDA plates as specified locations throughout the testing as per SOP.

→ Connect the Filtration manifold holder assembly with the S.S. reservoir properly with pipe and place sterilized S.S. cups in the sterile receptacle under Laminar air flow unit. Check the Manometer reading of working LAF and check the temperature as well as humidity of the sterility room Manometer reading of working LAF chamber pressure should be between 08 – 15mm of WG. Temperature reading of the sterility room should be 27ºC ± 2ºC.

→ Switch ON the vacuum pump but close the vacuum of the manifold holder assembly with the help of vacuum control key, present on the base of the individual manifold holder. Now place 0.45µm sterile membrane filters between filtration cup and receptacle with the help of sterilized forceps.

→ Wet the membrane filter by adding approx 15 ml of sterilized Fluid A (0.1% peptone water) to filter holder and filter the fluid by employing vacuum. Cut the tip of bottle/vial or ampoule with sterile SS blade in front of the gas burner and immediately transfer not less than half of the contents for LVP and the whole content of the vial for SVP to the membrane.

→ Immediately filter the solution with the aid of vacuum and wash the membrane three times with 100 ml of sterilized fluid A (Washing with fluid A is not required in case of sterile water for injection).

→ After complete filtration, stop the vacuum of the manifold with the help of manifold vacuum control key.

→ Lift the membrane carefully with the help of sterile forceps, aseptically cut the membrane filter into two halves with sterile SS scissor and transfer one half to FTM and one half to SCDM tubes by unplugging in front of gas burner only. Label both the tubes with the product name, B. No, lot No., date of testing, Completion date & Tested by.

→ Simultaneously prepare a negative control by filtering 100 ml of 0.1% peptone water instead of the product sample, cut the membrane into two halves with sterile SS scissor and transfer one half to FTM and one half to SCDM and label both the tubes as the Negative control.

→ Simultaneously prepare a chamber control during the sterility take two tubes, one is SCDM & other one is FTM tube, unplug the cotton plug of the tube and expose in LAF during sterility, after completion of sterility re-plug the tubes and then incubate the tubes as a chamber control.

→ After completion of work transfer all inoculated media through hatch box and then transfers all the equipment and exposed plates to microbiology analysis section.

→ Immediately move from the sterility area as per exit procedure specified SOP for Entry-Exit Procedure for Sterility Room.

→ Incubate the FTM tubes at 30ºC – 35ºC and SCDM tubes at 20ºC – 25ºC for 14 days.

→ The incubation period for terminally sterilized products is not less than 7 days and 14 days for aseptically filled products as per IP. If the Product is as per USP, BP, incubation period is 14 days for both terminally sterilized as well as for aseptically filled products.

→ Start the LAF of Microbiology Analysis room as per SOP and prepare four positive Control tubes by inoculating aseptically 10 to 100 cfu in FTM tubes with S. aureus NCIM 2079, P. aeruginosa NCIM 2200, B. subtilis NCIM 2063 and environmental flora EF - 1. Similarly prepare three SCDM positive control by inoculating approx 10 to 100 cells separately with C. albicans NCIM 3471, A. niger NCIM 1196 and environmental flora EF- 2. Incubate FTM positive control tubes at 30 – 35ºC for 3 days & SCDM positive control tubes at 20 – 25ºC for 5 days.

5.0 Observation and Interpretation of Results

⃝ Visually examine the media tubes daily to its conclusion for macroscopic evidence of microbial growth.

⃝ If no evidence of growth observed in any of the tube the product to be examined for the test complies with the test for sterility.

⃝ The test is not valid unless the negative control shows negative till at end of incubation, and positive control shows growth within specified incubation period.

⃝ Destroy all the positive controls after confirmation of growth in both FTM and SCDM tubes (Generally growth will be observed in 24 to 48 hrs).

⃝ If evidence of microbial growth is found in any of the tubes, investigate the cause of its failure as per SOP for sterility test failure investigation. If investigation report concludes that test is invalid, repeat the test with 20 units.

⃝ If no evidence of growth is found in the repeat test the product examined complies with the test for sterility. If evidence of microbial growth is found in the repeat test the product examined does not comply with the test for sterility.

Precaution

⃝ Sample bottles/ vials should be wiped properly with filtered 70% IPA before transferring them to sterility room and also inside the sterility room before sterility operations. Hatch box should also be cleaned properly with filtered 70% IPA solution.

⃝ Any material used in the sterility operation or inside sterility room should be autoclaved properly.

⃝ The tip of sample bottles/ vials should be heated properly in front of gas burner before cutting with a sterile heated cutting blade.

⃝ Clean the waste solution reservoir immediately after sterility operations and mop the outer surface of the reservoir with filtered 70% IPA with sterile mopper.

6.0 Frequencies

For LVP – Lot wise

For SVP – Batch wise

7.0 Abbreviation

SOP: Standard Operating Procedure

IPA: Isopropyl Alcohol

SVP: Small volume Parenterals

LVP: Large volume Parenterals

FTM: Fluid Thioglycollate Medium

SCDM: Soybean casein digest medium

Comments

Popular posts from this blog

Checklist for Audit in Sterile Area

Checklist for Audit in Sterile Area Checklist to verify the sterile manufacturing area before the audit. 1. Is there any SOP giving details for activity/ movement in the sterile area. 2. Verify the suitability of system for sanitization of hands. 3. Is there the SOP on entry-exit; gowning and de-gowning into the critical area available? 4. Verify the availability of following records:  Air pressure differential  Particulate monitoring  Temperature  Humidity  Microbiological monitoring by settle plate 5. Are these records available and maintained as per SOPs? 6. Check the above mentioned individual records. Are the entries within specifications mentioned in SOP/ MTPs. 7. Is the controlled copy of MTP pertaining to monitoring by settle plate available to plant person? 8. Are the location for settle plates identified and the chart/ layout available with plant? 9. Are the following areas monitored for parameters mentioned in step 4:  Air-lock to the critical ar...

Aseptic Filling Process (Media Fill) Validation Protocol in Sterile Pharmaceuticals

Aseptic Filling Process (Media Fill) Validation Protocol in Sterile Pharmaceuticals Validation of sterile manufacturing process by media fill validation test as per PICS guidelines for aseptic validation or aseptic process simulation. 1.0 OBJECTIVE 1.1 To define procedures for validating and maintaining the validation of all aseptic filling processes and qualification of the quality of the product by system/facility/equipment. 2.0 SCOPE 2.1 This procedure applies to all aseptically filled sterile products intended for human use. 3.0 RESPONSIBILITY 3.1 The Validation personnel coordinate the aseptic filling validation program and write the sterile media aseptic filling processes reports. 3.2 Manufacturing personnel the sterile media aseptic filling processes and performs the sterile media fills. 3.3 QA personnel perform the sampling and assist with the IPQA monitoring required for each sterile media fill. 3.4 QC personnel to perform the testing and assist with the monitoring required fo...

Blow Fill Seal (BFS) and Form Fill Seal (FFS) Technology in Sterile Manufacturing

Blow Fill Seal (BFS) and Form Fill Seal (FFS) Technology in Sterile Manufacturing Know about sterile pharmaceutical production using Blow Fill Seal (BFS) and Form Fill Seal (FFS) technology. Both of these techniques are used to manufacture  sterile pharmaceutical products   as parenteral (LVP & SVP), infusions, ophthalmic and inhalation products. These are automated techniques to prepare sterile products. The basic concept of the  FFS   and BFS is to reduce the  contamination   by forming the container, filling and sealing in a closed sterile chamber of the machine. There is no personnel intervention to reduce the chances of the contamination during the manufacturing of sterile products. It gives more production at very low operational cost with the high assurance of  sterility . Blow fill seal technology is widely used and accepted by the various pharmaceutical regulatory authorities as US-FDA and MHRA. The system is being used for over 30 years a...

MSDS (Material Safety Data Sheet)

MSDS   What Is an MSDS? An MSDS (Material Safety Data Sheet) is a document that provides detailed information about a chemical substance, including its hazards, safe handling procedures, storage requirements, and emergency measures. Today, MSDS is more commonly called SDS (Safety Data Sheet) under the Globally Harmonized System (GHS). Why MSDS / SDS Is Important Protect workers and employees Ensure safe handling and storage of chemicals Prevent accidents and injuries Provide emergency response information Comply with workplace safety regulations.  MSDS (Material Safety Data Sheet)  International safety standards.  GHS – Globally Harmonized System Full form: Globally Harmonized System of Classification and Labelling of Chemicals Develop by United Nations (UN).  OSHA (USA) – Hazard Communication Standard (HCS) Factories Act & Chemical Rules (India) Factories Act, 1948 Manufacture, Storage and Import of Hazardous Chemical Rules, 1989 REACH (Europe) REACH Regula...

Sterility Testing by Direct Inoculation Method

Sterility Testing by Direct Inoculation Method Direct Inoculation Method Learn the procedure for sterility test for sterile pharmaceutical products by direct inoculation method. 1.0 Objective :  To lay down Standard Operating Procedure is to provide guidelines for sterility testing by Direct Inoculation Method. 2.0 Scope This procedure is applicable for the products manufactured at the location which can not be tested by Membrane Filtration Method. 3.0 Responsibility Implementation: Microbiologist 4.0 Accountability Execution: Manager - QC Review & Approval : AGM - QA / QC Procedure Direct Inoculation Method Sample for Finished Products: Collect the samples to be tested for sterility. Out of the whole sample, randomly select 20 articles of each lot of batch for both LVP and SVP t erminal ly sterilized products and for aseptically filled products.    Sample for Intermediates: Randomly collect 16 pre-sterilized bottle samples to be tested for sterility from LVP bot...